ClearanceBrief
Recall Z-0128-2011 ยท FRN Pump, Infusion

Hospira Inc

Hospira brand Symbiq Two-Channel Infuser, LIST No. 16027 Symbiq Two-Channel Infuser, List Number Configurations: 16027-04-01/03/04, 51/52, 53/54, 79/80,81/82,83/84,87/88 and 89/90 16026-27-89/90, 16027-27-89/90, 16027-13-25/26,53/54, Distributed by: Hospira, Inc. Morgan Hill, CA

DateOct 22, 2010 (initiated Feb 22, 2010)
StatusTerminated
ReasonHospira has identified motor encoder failures in the Symbiq pumping mechanism that causes the infuser to cease operation during infusion therapy, resulting in delay or interruption of infusion therapy.
Root causeDevice Design
ActionHospira issued a Recall Notification letter dated October 4, 2010, updating the Urgent Device Field Correction letter dated February 22, 2010, ideentifying the affected products, the issue prompting the recall, and actions to be taken by customers. Customers are to examine their inventory to determine if they have any of the affected products, and recommended to remove them from service in critical care areas, and contact Hospira to schedule removal and replacement of the motor. Replace affected devices with a loaner or corrected back-up pump if available. Custoemrs can contacvtd Hospira Globa
DistributionWorldwide Distribution: USA and the coutnries of Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and the UAE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041550HOSPIRA PHOENIX INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE
Abbott Laboratories, Jun 30, 2004