ClearanceBrief
Recall Z-2503-2010 ยท GFA Blade, Saw, General & Plastic Surgery, Surgical

Linvatec Corp. dba ConMed Linvatec

GATOR. REF 9263A, 4.2mm STERLING, STERILE Rx Only.

DateSep 25, 2010 (initiated Aug 5, 2010)
StatusTerminated
ReasonConMed Linvatec, Largo, FL is recalling ConMed Linvatec 9263A,4.2mm Sterling Blade, Gator Meniscus Cutter, Lot Number 178026 for breech in the sterile barrier.
Root causePackaging process control
ActionRecall Notifications will be sent to all direct consignees by FedEx, email, Fax, etc. If the account no longer has the product in their possession, they will be asked to respond as such. A Return Material Authorizations Number will be assigned for the returned product as required. Product returned will be placed into quarantine and will be dispositioned in accordance with ConMed Linvatec procedures. 100% of the consignees will be contacted. The reply form is the mechanism to verify notification effectiveness to all consignees. If a reply is not received within approximately 30 business
Quantity174
DistributionProducts were distributed to ConMed Linvatec consignees, sales representatives and affiliates within the US and Australia. US distribution was to WA, AZ, PA, KS, NM, UT, ID, IL. Products were distributed to ConMed Linvatec Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.