Recall Z-0478-2011 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Sonosite, Inc.

The software is ARM version 60.80.101.025 (referred to as Nano 1.0.5). The software is used with the SonoSite NanoMaxx ultrasound system.

DateNov 26, 2010 (initiated Aug 13, 2010)
StatusTerminated
ReasonSoftware ARM version 60.80.101.025 (referred to as Nano1.0.5) on the NanoMaxx ultrasound system results in errors when the "print" command is selected. As a result, the displayed MI or TI values may be incorrect.
Root causeSoftware design
ActionSonosite issued Urgent Medical Device Correction letters dated August 9, 2010 to consignees. Consignees were advised of the software problem and given directions to verify the software version they are using. They are advised to contact SonoSite at 877-657-8118 to obtain a free software upgrade.
Quantity69
DistributionWorldwide Distribution: USA and to the countries of Australia, Great Britain, India, Israel, the Netherlands, New Zealand, Norway, and Spain.