Recall Z-0003-2011 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic

Verathon, Inc.

BVI 9600 Bladder Volume Instrument, Aorta Scan Mode The BladderScan BVI 9600 with AortaScan Mode is a user-selectable, dual-function ultrasound device that projects ultrasound energy either into the lower abdomen to obtain an image of the bladder for measuring bladder volume, or into the mid-abdomen to obtain an image of the abdominal aorta for aortic diameter measurements.

DateOct 4, 2010 (initiated Aug 23, 2010)
StatusTerminated
ReasonThe BladderScan BVI 9600 with AortaScan devices may experience the loss of a factory installed software configuration file which results in a loss of the aorta measurement function.
Root causeSoftware design
ActionVerathon sent an URGENT Medical Device Recall Notification letter dated August 12, 2010 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Verathon provided an addendum to the User Manual with instructions on how to reset the instrument. Customers were advised to: " Add the enclosed addendum to their current Users Manual. " Update the application software of the device to v2.7.0 or higher. If the customers had questions or needed support for the software update process, they were to contact Verathon Medical Customer Car
Quantity351 units
DistributionWorldwide Distribution - USA and the countries of Australia, Canada, China, Colombia, European Union, United Kingdom, Saudi Arabia, and Singapore.
FDA recordOpen on fda.gov