ClearanceBrief
Recall Z-1685-2011 ยท CGF Radioimmunoassay, Estrone

Beckman Coulter Inc.

Estrone Radioimmunoassay (RIA) Kit, DSL-8700. Intended for in vitro diagnostic use.

DateMar 17, 2011 (initiated Apr 22, 2010)
StatusTerminated
ReasonBeckman Coulter has confirmed that eight Estrone RIA kit (Ref DSL 8700) lots are inaccurately calibrated to the firm's internal reference calibrator. Lots 992637, 100115A, 100205A, 100205RA, and 100226A have now expired and should not be in active use any more.
Root causeOther
ActionBeckman Coulter released an Urgent: Product Corrective Action letter dated July 9, 2010 identifying the affected product and informing consignees of the reason for recall, asking them to take the following actions: *For lots 992637, 100115A, 100205A, 100205RA, and 100226A, review historical external quality control results relevant to the patients' results obtained with these lots to ensure the assay performance was within laboratory specifications. *For lots 100319A, 100409A, 100409RA, review patents' results obtained with the results of external quality control and interpret them in the lig
Quantity917 units
DistributionWorldwide Distribution -- US, including states of CA, FL, GA, IL, IN, KS, MI, NC, NY, OH, OR, TX, VA, WA, and WI and country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K935013DSL ESTRONE (DSL 8700)
Diagnostic Systems Laboratories, Inc., Jun 16, 1994