Recall Z-1374-2011 ยท LZS Excimer Laser System
AMO Manufacturing USA, LLC
AMO brand WaveScan WaveFront System, OPM software versions 3.67 and 3.67, Part Number: 0070-1531, Catalog Numbers: 0070-1664,0070-1649,0070-1654: Version 3.67 0070-1721: Version 3.671 0070-1662: Version 3.67, US version; 0070-1668; IVersion 3.67 International version Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The WaveScan WaveFront System is a class III
| Date | Feb 18, 2011 (initiated Jul 13, 2010) |
| Status | Terminated |
| Reason | The failure to comply with Minimum System Requirements, increased system errors could result, and a miscreated treatment table (incorrect LASIK refractive treatment) could result. |
| Root cause | Software design |
| Action | Abbott Medical Optics, Inc. sent an Urgent Medical Device Field Correction letter to all affected customers starting July 15, 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to take the following actions: Surgeons were instructed to complete the facsimile attachment to acknowledge receipt of the notification letter and fax to 714-247-4530. For those using the affected product a 3.68 OPM disc must be requested by e-mailing a request to OPM@pacful.com. For any questions please contact AMO Technical Service Representatiave at |
| Distribution | Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Great Britian, Greece, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Leb |
| FDA record | Open on fda.gov |