Recall Z-2201-2011 ยท LFX Enzyme Linked Immunoabsorbent Assay, Rubella

Bio-Rad Laboratories

Software CD is labeled in part: "BLOOD VIRUS DIVISION***BIO-RAD EVOLIS System APF Version 4.4 Software Version 2.00 2009/05***" The EVOLIS Assay Protocol Files (APF) Software contains assay-specific instructions necessary for the EVOLIS Microplate processing system to process enzyme immunoassays on the EVOLIS instrument, perform data reduction and generate reports.

DateMay 19, 2011 (initiated Jul 19, 2010)
StatusTerminated
ReasonThe EVOLIS Assay Protocol Files Software is configured in such a way that may result in a miscalculation of quantitative test results and/or an error in reporting positive/negative sample values.
Root causeSoftware design
ActionBio-Rad Laboratories sent an Urgent Product Correction letter dated July 19, 2010, to all customer sites via FedEx. There are two issues related to The EVOLIS Assay Protocol Files (APF) Software that could potentially affect patient results. Issue 1 involves quantitative sample test/calculations/results interpretation for assays with linear regression, point to point or 4PL data reduction. Issue 2 involves sample test results with " Result Report Flags''. A software patch to correct these two issues will be sent to each customer site. Bio-Rad Laboratories will work with each customer
Quantity272
DistributionNationwide Distribution -- Including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092353MONOLISA ANTI-HAV IGM EIA
Bio-Rad Laboratories, Inc., Oct 29, 2009
K961053SERAQUEST RUBELLA IGG
Quest Intl., Inc., Sep 4, 1996