Recall Z-1523-2011 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Siemens brand ARTISTE System Digital Linear Accelerator with the syngo RT Therapist version 4.1 Connected to MOSAIQ, Part No. 8168754, Distributed by Siemens Medical Solutions Oncology Care System, Concord CA Medical charged-particle radiation therapy system. The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-Ray photon and elec

DateMar 4, 2011 (initiated Jun 29, 2010)
StatusTerminated
ReasonA software problem that has the potential for data loss during the transfer of treatment records to the OIS, which may not be recorded in MOSAIQ, and subsequently may lead to mistreatment.
Root causeOther
ActionSiemens sent Urgent Medical Device Correction, Safety Advisory Notices to all affected customers on June 29, 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to include the Safety Advisory Notice in their System Owner Manual, Chapter, "Safety Advisory Letter. For questions customers were instructed to contact their Siemens Service Engineer or Application Specialist. For questions regarding this recall call 925-602-8157.
DistributionWorldwide Distribution - USA including NC,OH, and the countries of Belgium, Sweden, Germany, Lebanon, Japan, Australia, and Iran.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072485ARTISTE MV SA
Siemens Medical Solutions USA, Inc., Dec 27, 2007
K090683SYNGO RT THERAPIST CONNECT
Siemens Medical Solutions USA, Inc., May 8, 2009