Recall Z-2583-2010 ยท MQP Spinal Vertebral Body Replacement Device

Synthes USA (HQ), Inc.

SynFix-LR Spinal Implant System (Implant Holder for SynFix-LR and surgical technique brochure) The implant holder is used to place the implant into the vertebral disc space.

DateJul 23, 2010
StatusTerminated
ReasonA potential for the implant holder tip to break off and lodge in the SynFix-LR implant, which pose a potential risk to the patient.
Root causeError in labeling
ActionSynthes Spine Hospital sent an ' URGENT: MEDICAL DEVICE LABELING CORRECTION" letter dated August 20, 2010 to all customers. The letter described the labeling issue and actions to be taken by the customers. The customers were instructed to review the supplementary labeling provided; contact their Synthes Spine Sales Consultant to arrange for provision of the Holding Forceps (p/n 388.401) prior to each SynFix-LR procedure; and complete and return the enclosed Verification Section form to Synthes via fax at 610-251-9005 or email: FieldAction@Synthes.com. If you have any questions, please call 1
DistributionWorldwide Distribution: USA and the countries of Canada, Japan, Switzerland, Australia, Mexico, Malaysia, New Zealand, and Singapore.