Recall Z-2193-2010 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Stryker Endoscopy

Crossfire Console, model number 475-000-000, Manufactured by Stryker Endoscopy San Jose Medical device indicated for use in orthopedic and arthroscopic procedures for the knee, shoulder, ankle, elbow, wrist and hip. The system provides abrasion, resection, debridement, and removal of bone and soft tissue through its shaver blade and ablation and coagulation of soft tissue, as well as hemostasis o

DateAug 9, 2010 (initiated Jun 28, 2010)
StatusTerminated
ReasonRF energy or powered shaver handpieces may continue to operate when not intended when used with the iSwitch Wireless Foot switch.
Root causeSoftware design
ActionStryker Endoscopy sent an "Urgent: Device Correction" letter dated June 28, 2010, via FedEX Mailer to US customers. A letter dated July 9, 2010, was sent to International sites.The letter describes the product, problem, and action to be taken by customers. The Customers were instructed to follow the included instructions and use the USB dongle to install the correction and to complete and sign the enclosed "acknowledgement of receipt" form,upon completion of the software correction, and fax to (480) 754-8378 or scan in and email to crossfireconsole@stryker.com. The firm noted that-It is import
Quantity989 units
DistributionWorldwide distribution: USA and countries including Australia, France, Germany, Greece, South Africa, Iberia, Italy, New Zealand, Poland, UK, Switzerland, Canada, China, Japan and the EMEA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071859STRYKER CROSSFIRE SYSTEM
Stryker Endoscopy, Sep 20, 2007