ClearanceBrief
Recall Z-0117-2011 ยท NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)

Medtronic Inc. Cardiac Rhythm Disease Managment

Medtronic Virtuoso II DR, model D274DRG, and model D294DRG not available in the U.S.. Digital Dual chamber implantable cardioverter defibrillator (DDE-DDDR). Sterilized using ethylene oxide. Medtronic, Inc., Minneapolis, MN 55432 USA. Manufactured in: Juncos, Puerto Rico, USA. The Dual chamber implantable cardioverter defibrillator (ICD) is a multiprogrammable cardiac device that monitors a

DateOct 21, 2010 (initiated May 5, 2010)
StatusTerminated
ReasonMedtronic is informing you of a rare device software issue in certain ICD and CRT-D devices. A software update will be available to correct his issue. There have been 5 reports of this issue occurring with no reported patient injuries or deaths. Medtronic has identified the root cause to be a rare and specific sequence of events that must occur within a few milliseconds of each other. In the u
Root causeSoftware design
ActionMedtronic, Inc. sent an IMPORTANT MEDICAL DEVICE CORRECTION LETTER, dated May 2010 to physicians beginning May 05, 2010. The letter informed doctors of the issue and informed them that a software update would be available to correct this issue. The software was to be installed on the programmer, with a Medtronic representatives assistance, so that on a patient's next in-clinic interrogation the device would receive the upgrade automatically and their device would be corrected. For any questions regarding this recall physicians were to contact their local Medtronic Representative or Medtronic
Quantity13620
DistributionWorldwide Distribution - USA, Puerto Rico, Virgin Islands, and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussalam, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Gre
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.