Recall Z-2272-2010 ยท MAV Syringe, Balloon Inflation

Medtronic, Inc.

Medtronic Everest 30 Disposable Inflation Device, AC3200. Used to inflate/deflate balloon catheters as well as to monitor pressure within the balloon.

DateAug 25, 2010 (initiated Jul 16, 2010)
StatusTerminated
ReasonProduct sterility may be compromised due to breach of package seal integrity.
Root causePackaging process control
ActionMedtronic issued a Urgent Medical Device Recall Notice to US customer on 7/15/2010 and to foreign accounts advising of the issue and requesting that they segregate recalled product. A Medtronic Sales Rep will visit their facility to account for all affected product and initiate a credit. Foreign accounts will return product to their respective distribution centers.
Quantity768 units
DistributionWorldwide Distribution -- USA, including the state of MI and the countries of Austria, Belgium, Czech Republic, Finland, France, Germany, Greece, Ireland, Israel, Italy, Lithuania, Pakistan, Poland, Portugal, Slovakia, Spain, Sweden, Switzerland, Ukraine, United Arab Emirates, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K942269MEDTRONIC EVEREST INFLATION DEVICE
Medtronics Interventional Vascular, Oct 18, 1994
K960983MEDTRONIC EVERST 30 INFLATION DEVICE/SURVIVAL KIT
Medtronics Interventional Vascular, May 30, 1996