Recall Z-0067-2011 ยท LNX

SCC Soft Computer

SoftPath GUI Releases 3.17 through 4.1.0. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Manufacture/Distribution Dates: Version 3.17.5 10/14/2003. Version 3.17.6 03/28/2003. Version 3.17.7 07/02/2003. Version 3.17.8 03/03/2004. Version 4.1.0 07/03/2003. Version 4.1.1 10/03/2003. Intended for use in the Pathology Lab for receiving and accessioning specimens, documentin

DateOct 15, 2010 (initiated Mar 28, 2007)
StatusTerminated
ReasonSCC Soft Computer, Clearwater, FL issued a Correction for SoftPath GUI Releases 3.17 and 4.1 in March 2007 for clients with high volume output of final reports who use the Batch Print for Final Reports in the Scheduler, there is a potential for some reports not printing. This issue was corrected for the one affected user in October 2003.
Root causeSoftware design
ActionSCC Soft Computer sent a SAFETY NOTIFICATION TASK VERBIAGE dated January 5, 2004, to one client. The Notice identified the product, the problem, and the action to be taken by the client. SCC recommended the client set the Search Date Range to a small enough range to assure that the number of cases scanned during the job would be less than 10,000. The client was asked to acknowledge receipt of the task and to indicate if the Workflow or functionality was used/not used and Agree to use the alternative working solution. For any questions regarding this recall call (727) 780-0100, ext. 4235.
Quantity84
DistributionNationwide - USA, including the states of CA, CO, CT, FL, GA, IL, IN, LA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, PA, SD, TX, WI, and WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.