Recall Z-2263-2010 ยท BSZ Gas-Machine, Anesthesia

Philips Healthcare Inc.

Philips M1013A IntelliVue G1 Anesthesia Gas Modules with manifolds

DateAug 24, 2010 (initiated Jul 21, 2010)
StatusTerminated
ReasonThe manifold seal in the Philips M1657B Water Trap used on the Philips M1013A IntelliVue G1, M1019A, IntelliVue G5, and Anesthesia Gas Modules M1026B/BT may be unintentionally removed when changing the traps. A missing seal may result in leakage and may cause incorrect monitor readings of inspiratory or expiratory gas levels.
Root causeDevice Design
ActionPhilips Healthcare issued an "Urgent Device Correction" letter dated July 14, 2010 with an addendum to the Instructions for Use including additional instructions for the replacement of the water trap for the Philips M1013A IntelliVue G1, M1019A IntelliVue G5, M1026B/BT Anesthesia Gas Modules. This addendum is being sent along with the Urgent Medical Device Correction notice to all affected customers. Customers will be requested to incorporate the information contained in the Urgent Medical Device Correction notice and the Addendum to the Instructions for Use with the existing labeling (Instr
Quantity531 units
DistributionWorldwide Distribution: United States, and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Chile ,China ,Colombia, Czech Republic, Estonia, Finland, France, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Republic of
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060541THE PHILIPS MP20,MP30,MP40,MP50,MP60,MP70,MP80 AND MP90 INTELLIVUE ATIENT MONITORS
Philips Medizin Systeme Boblingen GmbH, Mar 24, 2006