Recall Z-2401-2010 ยท DWE Tubing, Pump, Cardiopulmonary Bypass

Maquet Inc.

Maquet HLM Tubing Pack; Bioline Coating

DateSep 12, 2010 (initiated Jul 13, 2010)
StatusTerminated
ReasonThe color of the tape on the arterial and the venous line were placed incorrectly, causing the connection of the arterial and the venous lines to be interchanged.
Root causeProcess control
ActionRecall notification letters were sent by Federal Express on July 13, 2010 to affected account and also an e-mail to the Maquet field representative who services the account.
Quantity5 units
DistributionDistribution: Pennsylvania
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053025JOSTRA HLM TUBING SET
Maquet Cardiopulmonary, AG, Nov 10, 2005