Recall Z-2240-2010 ยท LNH System, Nuclear Magnetic Resonance Imaging

Odin Medical Technologies Ltd.

PoleStar FlexiCoil band D/E, revisions A and B, Catalog # 9733063, Usage: Low field MRI device used to gather cranial images.

DateAug 19, 2010 (initiated Jul 1, 2010)
StatusTerminated
ReasonElectronic box on headstrap may overheat, causing burn to patient.
Root causeDevice Design
ActionCustomers were notified by letter on 07/07/2010, advising them of the overheating problem and alerting them that a service rep will be visiting to inspect and upgrade any affected FlexiCoils. Questions were referred to 800-595-9709.
Quantity560 units
DistributionCA, CO, DC, IL, IN, MD, MI, MN, MO, NE, NJ, NV, NY, OH, OR, WA, and WI. Foreign distribution to Austria, Belgium, Canada, China, France, Germany, Israel, Italy, Korea, Netherlands, Poland, Spain, Russia, Switzerland and Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032541POLESTAR N-20
Odin Medical Technologies , Ltd., Sep 12, 2003