Recall Z-2222-2010 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Siemens Medical Solutions USA, Inc

SIEMENS AXIOM Luminos TF and Sireskop SD Solid State x-ray imager (flat panel/digital imager), Fluoroscopic x-ray diagnostic system.

DateAug 13, 2010 (initiated Jul 13, 2010)
StatusTerminated
ReasonFluid may enter the multi-functional handle (Opti Grip) when using the Axiom Luminos TF or Sireskop SD system, causing unintended reaction or movement, increased radiation or useless images.
Root causeDevice Design
ActionSiemens Healthcare Sector sent an "Urgent Field Safety Notice" to consignees describing the issue and informing them to take specific precautions when performing interventional procedures to avoid fluids penetrating the system. The multi-functional handle (Opti Grip) will be replaced through implementation of Update Instructions AX030/10/S. The modified version of the handle resists fluids from entering the handle and prevents malfunction.
Quantity1296
DistributionNationwide distribution: all 50 US states and PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051602AXIOM SIRESKOP SD
Siemens Medical Systems, Inc., Jul 7, 2005
K062623AXIOM LUMINOS DRF
Siemens Medical Solutions USA, Inc., Aug 22, 2007