Recall Z-0322-2011 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips Healthcare Inc.

HeartStart MRx Defibrillator/Monitor Models: M3535A / M3536A / M3536J / M3536MC The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on th

DateNov 8, 2010 (initiated Jul 16, 2010)
StatusTerminated
ReasonUpdated Instructions for Use- HeartStart MRx monitor/defibrillator. The update provides more detailed information on: " Default ECG Lead Behavior " Synchronized Cardioversion using External Paddles
Root causeLabeling design
ActionPhilips issued an Urgent Voluntary Medical Device Correction notification/Field Safety Notice and an Instructions for Use Addendum on July 20, 2010, to affected customers. The Notice identified the product, the problem, and the action customers should take. In the USA, the notification sent to all affected customers via UPS and confirmation of delivery would be tracked. Outside the USA, Philips representatives in the affected geographies would manage their mailings locally. Customers were instructed to review the HeartStart MRx Instructions for Use addendum and keep a copy with their equipm
Quantity41,119 units
DistributionWorldwide Distribution - US Nationwide and the countries of: CANADA AFGHANISTAN ALBANIA ARGENTINA ARUBA AUSTRALIA AUSTRIA AZERBAIJAN BAHRAIN BELGIUM BOSNIA&HERZEGOVINA BRAZIL BRUNEI DARUSSALAM BULGARIA CHILE CHINA COLOMBIA CYPRUS CZECH REPUBLIC DENMARK ECUADOR EGYPT ESTONIA FINLAND FRANCE GERMANY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051137TECHDEVICE OCCLUSION BALLOON CATHETER
Techdevice Corporation, Feb 24, 2006