Recall Z-2653-2010 ยท DEW Igg, Antigen, Antiserum, Control

Beckman Coulter Inc.

SYNCHRON System(s) Immunoglobulin A (Ig-A) Reagent, Part Number: 467920

DateSep 30, 2010 (initiated Jun 28, 2010)
StatusTerminated
ReasonThe recall was initiated after Beckman Coulter Inc. confirmed reports of high recovery of immunoglobulin A (IgA) in proficiency surveys or patient samples using Synchron Ig-A reagent lots M812164 or M902345. Using these reagent lots, some samples deficient in IgA have yielded results in the normal range. Other samples with normal levels of IgA have yielded falsely high results.
Root causeOther
ActionA Product Corrective Action (PCA) letter was sent on the week of June 28, 2010 to the affected customers. The letter provided the customers with an explanation of the problem identified and advised customers to discontinue use of the affected lots of the SYNCHRON System(s) Immunoglobulin A (Ig-A) Reagent. Customers were instructed to: (1) Discontinue use of Ig-A reagent lots M812164 and M902345. Contact your Customer Service Representative at (800) 526-3821 in the United States or (800) 463-7828 in Canada or contact your local Beckman Coulter Representative to discuss reagent replacement opt
Quantity852 units
DistributionUS Nationwide & Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K965108SYNCHRON LX SYSTEMS IMMUNOGLOBULIN G/A/M REAGENT
Beckman Instruments, Inc., Mar 6, 1997