ClearanceBrief
Recall Z-0438-2011 ยท KYI Prosthesis, Wrist, Carpal Trapezium

Ascension Orthopedics, Inc

The Ascension PyroHemiSphere (aka NuGrip), Model #40-M NUG-443-40M, Catalog #NUG-443-40M, Lot #10-0469, is a single-use uncemented, one-component prosthesis for the basal thumb joint. The NuGrip is intended to replace the proximal end of the first metacarpal.

DateNov 22, 2010 (initiated Jul 12, 2010)
StatusTerminated
ReasonTwo different product lots were found to be incorrectly labeled and placed into the incorrect box.
Root causeMixed-up of materials/components
ActionAscension Orthopedics notified by phone on June 25, 2010 and sent a letter dated July 22, 2010, to all customers. The customers were instructed to return the devices to the firm. Should you have any questions or concerns please feel free to contact (512) 836-5001 ext 1585.
Quantity10
DistributionWorldwide distribution: USA including states of: IL, IN, MD, OH, and TN; and countries of : Puerto Rico and Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041451ASCENSION PYROHEMISPHERE, MODELS PHS-440-10, PHS-440-20, PHS-440-30, PHS-440-40, PHS-440-50
Ascension Orthopedics, Inc., Aug 25, 2004