ClearanceBrief
Recall Z-2199-2010 ยท HTY Pin, Fixation, Smooth

Smith & Nephew Inc

PERI-LOC, 2.0 MM X 228 MM, K-WIRE W/TROCAR POINT, CO-CR, REF 71173361, QTY: (6), STERILE FR, CE 0123, 2015/04, Smith & Nephew, Inc., Memphis, TN 38116, EC REP: Smith & Nephew Orthopaedics GmbH, Tuttlingen, Germany Usage: Bone fracture fixation

DateAug 10, 2010 (initiated Jun 28, 2010)
StatusTerminated
ReasonPlastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, compromising sterility.
Root causeProcess control
ActionAll affected Smith & Nephew Sales Representatives were notified of problem and the recall was sent via overnight mail on 01/23/2009. All affected Smith & Nephew International Distributors were notified via email and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return.
Quantity8 units
DistributionKY, TX, DC, NC, CA, AZ, GA and OH; 3 International: Australia, Malaysia, and Switzerland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.