ClearanceBrief
Recall Z-2446-2010 ยท IKT Electrode, Needle, Diagnostic Electromyograph

Chalgren Enterprises Inc

Disposable Detachable Monopolar Needle Electrode for electromyograph model number 110-725, for Physician use, manufactured by Chalgren Enterprises Inc., Gilroy, CA.

DateSep 21, 2010 (initiated Apr 12, 2007)
StatusTerminated
ReasonSterility Compromised-- Investigation found that individual packaging seals were separating, which may compromise sterility.
Root causePackaging
ActionChalgren Enterprises, Inc. removed all of the affected inventory from inventory and had all affected inventory retrieved from distribution.
DistributionNationwide Distribution: USA states of CA, IL, NJ, OH and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K912283DETACHABLE, DISPOSABLE MONOPOLAR NEEDLE ELECTRODE