Recall Z-2192-2010 ยท HRX Arthroscope
Linvatec Corp. dba ConMed Linvatec
REF 87100 87k Arthroscopy Tubing Set. For use with the 87k Arthroscopy Pump (87000). STERILE EO. Rx ONLY. Warning: Do not use this tubing set with any Small Joint Cannula System. CONMED LINVATEC, 11311 CONCEPT BLVD, LARGO, FL 33773-4908 USA. Made in USA. The 87K Arthroscopy Tubing Set is designed for use with the 87K Arthroscopy Pump System. This system can be used in arthroscopic procedure
| Date | Aug 9, 2010 (initiated Jul 6, 2010) |
| Status | Terminated |
| Reason | ConMed Linvatec Largo, FL is recalling 87K Arthroscopy Tubing Set, Product Number 87100, Lots 0904151 to and including 1005201. Product may have a breach in the seal that could potentially compromise the sterility of the contents. |
| Root cause | Process control |
| Action | ConMed Linvatec Corporation sent an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" letter dated July 6, 2010, via FedEx next day delivery to all customers. The letter described the product, problem and action to be taken by customers within 30 days of receipt of the letter. The customers were asked to PLEASE DISCONTINUE USE OF THE IDENTIFIED PRODUCT WITH LOT NUMBERS BEGINNING WITH 0904151 THROUGH AND INCLUDING 1005201; to immediately check their inventory for the affected lots number and Do Not Use if they have any of the products; to complete and return a Reply Form and to notify their custom |
| Quantity | 97,830 |
| Distribution | Worldwide distribution: USA and Canada. |
| FDA record | Open on fda.gov |