Recall Z-0457-2011 ยท MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator

Respironics California Inc

Respironics V60 Ventilator, manufactured by Respironics California Inc. in Carlsbad, CA.

DateNov 24, 2010 (initiated Jul 2, 2010)
StatusTerminated
ReasonRespironics California Inc has received customer complaints of an odor or burning smell coming from the V60 Ventilator, and identified a component on the Power Management PCB that exhibited thermal damage.
Root causeDevice Design
ActionPhililps Healthcare issued an Urgent Medical Device Field Correction letter dated July 2, 2010 concerning the RespironicsV60 Ventilator issue. Customers will be contacted by US Field Service Technicians and international distributors to schedule and perform inspections and replacement of the affected component. The unit may continue to be used pending completion of the field inspection. Customers may contact Respironics US Customer Service about this action at 1 800 345 6443, option #5, then option #1.
Quantity504
DistributionWorldwide Distribution: Throughout the US and Puerto Rico,, and to the countries of: Australia, Austria, Belgium, Canada, Denmark, France, Germany, Hong Kong, India, Italy, Kenya, Kuwait, Libya, Malaysia, Oman, Netherlands, Norway, the United Kingdom, Saudi Arabia, and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082660V60 VENTILATOR, MODEL V8000
Respironics California, Inc., Jan 9, 2009