Recall Z-2449-2010 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems North America Co. Phillips

BIPLANE X-ray Imaging Systems included Allura XPER FD10/10, FD20/10, FD20/20, Integris BH5000, and Integris Allura 12-15 Biplane. The BIPLANE system has two C-arms and it allows the operator to image in two planes: Lateral and Frontal. It has a C-arm on the stand and C-arm suspended from the ceiling. The product's brochure is labeled in part: "Philips Healthcare Global Information Center...P

DateSep 22, 2010 (initiated Feb 4, 2010)
StatusTerminated
ReasonA lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.
Root causeSoftware design
ActionOn 02/04/2010, Philips Healthcare began sending out the URGENT-Device Correction Notices to their consignees. The firm issued two notifications, one for the Allura Integris and Xper FD10C, FD10F, FD10F/C and FD20 Monoplane X-Ray imaging systems with the Velara Generator. The second notification is for the Allura Integris Xper FD10 and FD20 Biplane X-ray imaging systems with the Velara Generator. This notice informs the consignees that a system locks up during a critical interventional case and it might impose a serious injury to patients. This lock up issue could only be resolved by a warm
Quantity267
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031333PHILIPS INTEGRIS ALLURA FLAT DETECTOR RELEASE 1.2
Philips Medical Systems North America, Inc., May 13, 2003
K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003
K984545RELEASE 2 OF PHILIPS INTEGRIS SERIES (V5000, BV5000, H5000, BH5000, HM2000) SYSTEMS
Philips Medical Systems North America, Inc., Feb 25, 1999