Recall Z-2448-2010 ยท IZI System, X-Ray, Angiographic
Philips Medical Systems North America Co. Phillips
MONOPLANE X-ray Imaging Systems include the Allura XPER FD10 F/C, Allura XPER FD20, Integris H5000/Allura 9 F/C, Integris CV, and Allura 12-15 Monoplane. The Monoplane system has one C-arm and it could be mounted on a ceiling or on the floor stand. The product's brochure is labeled in part: "Philips Healthcare Global Information Center...P.O. Box 1286...5602 BG Eindhoven...The Netherlands".
| Date | Sep 22, 2010 (initiated Feb 4, 2010) |
| Status | Terminated |
| Reason | A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem could only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted. |
| Root cause | Software design |
| Action | On 02/04/2010, Philips Healthcare began sending out the URGENT-Device Correction Notices to their consignees. The firm issued two notifications, one for the Allura Integris and Xper FD10C, FD10F, FD10F/C and FD20 Monoplane X-Ray imaging systems with the Velara Generator. The second notification is for the Allura Integris Xper FD10 and FD20 Biplane X-ray imaging systems with the Velara Generator. This notice informs the consignees that a system locks up during a critical interventional case and it might impose a serious injury to patients. This lock up issue could only be resolved by a warm |
| Quantity | 1470 units |
| Distribution | Nationwide Distribution |
| FDA record | Open on fda.gov |