Recall Z-2448-2010 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems North America Co. Phillips

MONOPLANE X-ray Imaging Systems include the Allura XPER FD10 F/C, Allura XPER FD20, Integris H5000/Allura 9 F/C, Integris CV, and Allura 12-15 Monoplane. The Monoplane system has one C-arm and it could be mounted on a ceiling or on the floor stand. The product's brochure is labeled in part: "Philips Healthcare Global Information Center...P.O. Box 1286...5602 BG Eindhoven...The Netherlands".

DateSep 22, 2010 (initiated Feb 4, 2010)
StatusTerminated
ReasonA lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem could only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.
Root causeSoftware design
ActionOn 02/04/2010, Philips Healthcare began sending out the URGENT-Device Correction Notices to their consignees. The firm issued two notifications, one for the Allura Integris and Xper FD10C, FD10F, FD10F/C and FD20 Monoplane X-Ray imaging systems with the Velara Generator. The second notification is for the Allura Integris Xper FD10 and FD20 Biplane X-ray imaging systems with the Velara Generator. This notice informs the consignees that a system locks up during a critical interventional case and it might impose a serious injury to patients. This lock up issue could only be resolved by a warm
Quantity1470 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022899PHILIPS INTEGRIS ALLURA FLAT DETECTOR RELEASE 1.0
Philips Medical Systems North America Co., Nov 22, 2002
K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003
K984545RELEASE 2 OF PHILIPS INTEGRIS SERIES (V5000, BV5000, H5000, BH5000, HM2000) SYSTEMS
Philips Medical Systems North America, Inc., Feb 25, 1999