Recall Z-2201-2010 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips Healthcare Inc.

Philips Intellivue Patient Monitor Model MP60 (M8005A)

DateAug 10, 2010 (initiated Jun 28, 2010)
StatusTerminated
ReasonAdditional Instruction for Use for Intellivue MP60/MP70 to prevent monitors from becoming loose and detaching from the mounting arm.
Root causeLabeling design
ActionPhilips issued on 6/28/10 an Urgent Medical Device Correction notification/Field Safety Notice. Included with the Urgent Medical Device Correction notification/Field Safety Notice is an updated instructions for use and service instructions for these monitors that remind users to follow the mounting arm manufacturer's instructions for use when repositioning the monitor and instructing them to perform a periodic check that the monitor is not loose on the arm.
Quantity11,529 units
DistributionWorldwide distribution: USA, Canada, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, Bosnia, BRAZIL, BULGARIA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HU
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093268INTELLIVUE PATIENT MONITOR
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN, May 11, 2010