Recall Z-2111-2010 ยท GDF Guide, Needle, Surgical
Carefusion 2200 Inc
Temno Coaxial Chiba Fine Needle Aspiration; a coaxial introducer needle (20G x 15 cm) with a Chiba biopsy needle (22G x 20 cm); sterile, for single use only sets, 5 sets per case; CareFusion, McGaw Park, IL 60085 USA, Made in Dominican Republic; REF CCH2220 For use in soft tissue aspiration such as breast, kidney, liver, lung, thyroid, lymph nodes and other various soft tissue masses. It is not i
| Date | Jul 28, 2010 (initiated Jun 25, 2010) |
| Status | Terminated |
| Reason | The introducer needle is only 10 cm in length instead of the required 15 cm. |
| Root cause | Mixed-up of materials/components |
| Action | Cardinal Health, on behalf of CareFusion, telephoned the customers on June 15, 2010 and sent a follow-up "URGENT: PRODUCT RECALL" letter on the same date via UPS next day air. The letter describes product, problem, and action taken by customers. The customers were instructed to immediately stop using affected product, return the enclosed acknowledgment form via fax to 847-686-9101 (note: Cardinal Health is required to confirm receipt of this notification from the customers), notify any customers to whom they may have distributed product to, contact CareFusion Customer Service at 800-653-6827 |
| Quantity | 65 units |
| Distribution | Nationwide distribution: California, Florida, Georgia, Illinois, Kentucky, Massachusetts, New Jersey, New York and Pennsylvania. |
| FDA record | Open on fda.gov |