Recall Z-0429-2011 ยท NOE Test, System, Immunoassay, Lipoprotein-Associated Phospholipase A2

DiaDexus, Inc

diaDexus PLAC Test ELISA Kit; Manufactured and distributed by diaDexus , Inc., South San Francisco, CA; Model Part Number: 90123 The diaDexusPLAC Test ELISA Kit is an enzyme Immunoassay for the quantitative determination of Lp-PLA2 (lipoprotein-associated phospholipase A2) in human plasma and serum, to be used in conjunction with clinical evaluation and patient risk assessment as an aid in predict

DateNov 22, 2010 (initiated Mar 15, 2010)
StatusTerminated
ReasonThe reason for the recall is that Real Time Stability Test program showed the kit lot failed stability prior to expiration, affecting samples with results> 500 ng/ml. Results up to 500 ng/ml exhibit passing stability results and are not impacted by this issue.
Root causeOther
ActionDiaDexus notified all customers by phone, including distributor's customers, beginning on March 15, 2010. Customers were told to re-test samples that were tested from 2/1/2010 through 2/28/2010 which have values above 500 ng/ml with a new lot of product. For questions regarding this recall call (650) 246-6477.
Quantity239 kits
DistributionNationwide Distribution - US including NJ, OH, KS, IN, FL, CA, UT, OK, PA, WA and MI.