Recall Z-0455-2011 ยท MUJ System, Planning, Radiation Therapy Treatment

Varian Medical Systems Oncology Systems

Varian Medical Systems, Inc. Eclipse/VARiS Vision, an Eclipse radiotherapy Treatment Planning System (Eclipse TPS); Eclipse Version 7.3.10 (multiple workstations) VARiS Vision System 7. x; Model Numbers: H48/H83.

DateNov 24, 2010 (initiated May 13, 2010)
StatusTerminated
ReasonIncorrect treatment plan or record: A software design deficiency has the potential for corrupted treatment plan records or treatment records in cases where multiple users editing a record simultaneously can result in a corrupted (but accessible) treatment plan record.
Root causeSoftware design
ActionOn May 13, 2010, the firm issued letters to all its direct consignees, informing them of the affected products and providing instructions on the recall. ***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
Quantity2417 units***12-20-12 AMENDED TO: 49 units***
DistributionNationwide and Worldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071873MODIFICATION TO ECLIPSE TREATMENT PLANNING SYSTEM
Varian Medical Systems, Inc., Aug 9, 2007
K932501VARIAN ORIGAMI, VERSION 1.0
Varian Assoc., Inc., Sep 23, 1994