ClearanceBrief
Recall Z-0395-2011 ยท IXQ Table, Radiographic, Stationary Top

Med Tec Inc

CIVCO Type-S Extension, REF MTIL6605, Reusable non-sterile extension for use with Universal Couchtop. The firm name on the label is CIVCO, Orange City, IA. For use to support and aid in positioning a patient during radiologic and other medical procedures.

DateNov 16, 2010 (initiated Jun 21, 2010)
StatusTerminated
ReasonUnder certain conditions, the extensions can become detached from the main body of the couch.
Root causeDevice Design
ActionThe recalling firm, CIVCO, issued an 'Urgent Field Safety Notice" dated June 15, 2010, via overnight mail on June 21, 2010, regarding the first issue for the three extension models. The letter described the product, problem, and actions to be taken by the customers. The customers were instructed to load the patients on the couch from the bottom of the couch and not from the superior end with the extension and do not let patients crawl to the top of the couch with their full upper body weight at the end of the couch; notify others within their organization that may have the extension, and com
Quantity3 extensions
DistributionWorldwide distribution: USA including states of: CA, FL, IA OH, and WI; and countires of: United Kingdom, Australia, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973842CARBON FIBER CONFORMAL COUCH TOP
Medtec, Inc., Dec 5, 1997