ClearanceBrief
Recall Z-0037-2011 ยท LDS Device, Pasteurization, Hot Water

Cenorin

HLD SYSTEM 540 is the Medical Device Cleaning and High Level Disinfection Systems (Washer/Pasteurizers). This device does not have the external heater unit known as High Throughput Heater. Each unit has a metal serial tag that is riveted to the enclosure of the unit and it is labeled in part: "HLD SYSTEMS 540***FOR SALES.SERVICE.SUPPLIES CALL***Cenorin*** Kent, WA 98032***P:800.426.1042***F: 253

DateOct 12, 2010 (initiated Jun 7, 2010)
StatusTerminated
ReasonThe water level sensor switch in the Medical Device Cleaning and High Level Disinfection Systems (Washer/Pasteurizers) is sensitive to externally induced vibration noise, and it produces a false signal that the water level in the tank is full while it is only 75% full. It could present a health risk of biological hazard because the medical devices are not being adequately cleaned or pasteurized
Root causeComponent design/selection
ActionOn June 7, 2010, firm, CENORIN, sent "URGENT: MEDICAL DEVICE RECALL" Initial Notification letter with Recall Response Form, Series 500 Water Level Check Instructions, and the Water Level Switch Upgrade Kit Instruction (service kit P/N # 082395-00) to their consignees. The firm also sent these documents in Spanish language to their foreign consignees. The letter describes the recall product as HLD Systems Medical Device Cleaning and High Level Disinfection Series 500 Washer/Pasteurizer-Model 540, Model 540HT and Model 520. Customers are advised to call the Customer Service at 1-800-426-140
Quantity16 units
DistributionWorldwide distribution:USA and countries including: Costa Rica, Ecuador, Japan, Mexico, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.