Recall Z-2178-2011 ยท JIT Calibrator, Secondary

Beckman Coulter Inc.

Beckman Coulter Access Free T3 Calibrators, Part Number A13430 in vitro diagnostic.

DateMay 10, 2011 (initiated Dec 17, 2009)
StatusTerminated
ReasonBeckman Coulter has confirmed reports that the Access Free T3 calibrator lot numbers 989902, 989903, 989904, 989905, 989906 and 989907 will show a decline in performance before reaching their stated expiration date. The affected calibrator lots will cause one or more quality control samples to produce out-of-range high result.
Root causeNonconforming Material/Component
ActionBeckman Coulter sent a letter dated December 29, 2009, providing the customers with an explanation of the reason for recall and asked them to take the following actions: - Discontinue use of the lots of reagent listed above. - Review your historical quality control to ensure that the assay performance of these lots has been within your laboratory specifications. NOTE: Out-of-range high quality control results may indicate invalid test results. Consignees were asked to share this information with their laboratory staff and retain this notification as part of their laboratory Quality System do
Quantity2,320 kits
DistributionDistribution Nationwide and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042651FREE T3 CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS
Beckman Coulter, Inc., Oct 5, 2004