Recall Z-1956-2010 ยท DZL Screw, Fixation, Intraosseous

Biomet 3i, LLC

BIOMET 3i GS300 GOLD SLOTTED SCREW-3mm, STERILE, Rx Only. Made in USA. Lots 868517, 865348, and 868515. BIOMET 3i, 4555 Riverside Dr., Palm Beach Gardens, FL.

DateJul 6, 2010 (initiated Oct 6, 2009)
StatusTerminated
ReasonBiomet 3i, Palm Beach Gardens, FL is recalling their Gold Slotted Retaining Screw 3MM, Model Number GS300. Product is Out Of Specification - product was manufactured with a material not specified for use in this product.
Root causeOther
ActionEach US customer was contacted by telephone using a prepared phone script. Following telephone contact, Biomet 3i performed a follow-up of each call with a faxed letter. International customers were first notified by e-mail then by fax letter. Each customer was instructed to check their respective inventory for the affected product and corresponding lot number. They were then requested to fax back a response form to Biomet 3i Regulatory Services to provide feedback concerning their inventory. They were also instructed to return the affected product to Biomet 3i. Questions or concerns a
Quantity495 screws
DistributionProduct was distributed to 3 distributors, 14 direct consumer/users. Product was distributed in Japan, Spain, Switzerland, Uruguay, Paraguay, Mexico, France, Netherlands, Australia, and in the US in KY, DC, IL, RI, MA, NJ and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K874400NON-ROTATING GOLD CYLINDER AND SCREW
Implant Innovations International, Inc., Mar 2, 1988