Recall Z-2206-2010 ยท DZL Screw, Fixation, Intraosseous

Biomet Microfixation, Inc.

BIOMET MICROFIXATION "1.5MM" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW, QTY 1, 1.5 X 4MM TITANIUM. REF 91-1504 1 COUNT and REF 95-1504 - 5 PACK. BIOMET MICROFIXATION 1520 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA, WWW.BIOMETMICROFIXATION.COM. Intended to aid in the alignment and stabilization of bone in the oral cranio-maxillofacial skeletal system.

DateAug 11, 2010 (initiated Mar 17, 2010)
StatusTerminated
ReasonBiomet Microfixation has received a complaint from a customer who reports finding a 3.5mm length screw in one 5 pack of 4mm length screws.
Root causeOther
ActionBiomet Microfixation is recalling their 1.5x4mm HT X-drive Screws Lot Number 148530 after receiving a customer reported finding one 3.5MM screw in a package labeled as containing 4.0MM screws. Lot 148530 was packaged into one count packages 91-1504 and 5 count packages 95-1504. The 1 and 5 count packages were distributed between 12/10/2009 and 02/01/2010. Beginning on 03/17/2010 customers identified as receiving Lot 148530 HT X-drive Screws were notified of the recall by a combination of an Urgent Medical Device Recall Notice letter, e-mail, fax and telephone calls. Customers were instru
Quantity1247 screws
DistributionWorldwide Distribution -- USA, including states of FL, GA, IN, LA, MD, MI, MO, NC, NY, OK, TX, and VA and countries of Philippines, Japan, Brazil, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953385LORENZ 1.0MM, 1.5MM, 2.0MM SYSTEM
Walter Lorenz Surgical, Inc., Sep 28, 1995