Recall Z-2274-2010 ยท KGZ Accessories, Catheter

Arrow International Inc

Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr. Catheters; Catalog number AJ-09801

DateAug 25, 2010 (initiated Jun 7, 2010)
StatusTerminated
ReasonSterility not assured. The stopcock handle position was inadvertently changed from an open to a closed position, which prevents sterilization of the closed portion of the device. Sterility of the affected product cannot be guaranteed.
Root causeProcess change control
ActionJune 7, 2010 Notification letter to customer instructed them to return the product to Arrow. Arrow will destroy the returned product.
Quantity3000 kits
DistributionAll affected product was distributed to one foreign consignee: Arrow Japan Ltd.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K781846ADAPTER, CATHETER SHEATH
Arrow Intl., Inc., Dec 12, 1978