Recall Z-2420-2010 ยท HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment

Stryker Instruments Div. of Stryker Corporation

Stryker Bur Guard REF 2296-301, Stryker Instruments, Kalamazoo, MI (USA) The Impaction Bur Guard is intended to be used with the Impaction Drill. The drill is used for surgical procedures involving precision drilling and cutting of bone and hard tissue. The bur guard covers the nose of the handpiece and the shaft of the cutting accessory to prevent exposure to heat and moving parts, as well as min

DateSep 13, 2010 (initiated Jun 2, 2010)
StatusTerminated
ReasonThe reason for the field action is that the inner diameter of the Bur Guards may be out of specification. If there is excess material on the inside of the Bur Guard at the location where the guard tapers, it may come into contact with the Nose Cone of the drill. This could result in an injury or burn to the patient and/or user.
Root causeDevice Design
ActionStryker Instruments sent an Urgent Medical Device Recall Notification dated June 10, 2010, to their customers. The letter described the product the problem and the action to be taken by the customers. The customers were instructed to: 1) immediately review this recall notification. 2) forward this letter and the attached Business Reply Form (BRF) to all affected locations. 3) Immediately check and quarantine any affected product found. 4) Destroy all affected product by breaking off at least 3 of the tabs on the bur guard and then dispose of the affected product in normal trash. 5) Co
Quantity10982
DistributionWorldwide Distribution: US AND COUNTRIES OF SCANDINAVIA, AUSTRALIA, CANADA, CHILE, CHINA, BRAZIL, FRANCE, GERMANY, GREECE, SOUTH AFRICA, ITALY, JAPAN, MEXICO, SWITZERLAND, PORTUGAL, SINGAPORE and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961970COMMAND 2 MICOELECTRIC SURGICAL SYSTEM
Stryker Corp., Jul 26, 1996