Recall Z-2238-2010 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

GE Healthcare, LLC

Innova 2000, manufactured by GE Medical Systems, SCS

DateAug 19, 2010 (initiated Jun 10, 2010)
StatusTerminated
ReasonA hardware component (potentiometer and encoder)mismatch in Innova 2000 digital fluoroscopic imaging systems may cause a Source to Image Distance (SID) failure, which will cause the imager to fail to detect the distance between the imager and the patient. This will result in loss of the x-ray image if the recovery process is not performed. The error message to use the Emergency Back-Out Button for
Root causeOther
ActionGE Healthcare sent a GE "Urgent Medical Device Correction" dated June 9, 2010letter to consignees. The letter was addressed to Hospital Administrators / Risk Mangers; Mangers of Radiology/Cardiology and Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product Details; Safety Instruction; Product Correction and Contact Information. The letter detailed the method for identifying SID failure and the use of the emergency back-out buttoms to recover the X-ray image. A GE service representative will affix a label to the device on-site to remind operators how to recover
Quantity885
DistributionWorldwide distribution: States of AL, AZ, AR, CA, CO, CT,DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NV,NH, NJ, NM, NH,NC, ND, OH, OK OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, and WI, and countries of VENEZUELA, UNITED KINGDOM, TURKEY, TUNISIA, THAILAND, TAIWAN, SYR
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022322INNOVA AND INNOVA 2000 S
Ge Medical Systems, Inc., Oct 8, 2002
K024200INNOVA 2000 AND INNOVA 2000S (MOBILE VERSION)
Ge Medical Systems, Inc., Jan 16, 2003
K993037LCV+ VERSION 2
Ge Medical Systems, Inc., Feb 14, 2000