Recall Z-2025-2010 ยท GNC Antigens, Febrile, Slide And Tube, All Groups, Salmonella Spp.
Becton Dickinson & Co.
Difco Salmonella 0 Group A Antigen, catalog #240731, packaged in 5 mL glass vials, labeled in part ***Becton, Dickinson and Company, Sparks, MD 21152***
| Date | Jul 15, 2010 (initiated May 19, 2010) |
| Status | Terminated |
| Reason | In vitro diagnostic reagent to aid in the diagnosis of Salmonellosis may exhibit decreased or no reactivity. If proper controls are run with the lot in question, it would become apparent that the antigen is non reactive and patient specimen results should not be reported. If control testing is not performed a salmonella-infected patient could go undiagnosed, with possible adverse effects. |
| Root cause | Nonconforming Material/Component |
| Action | The recalling firm notified customers by fax and e-mail on 05/19/10. The notification advised that the referenced lot number of Salmonella 0 Group A Antigen may exhibit decreased or no reactivity when testing the patient's serum directly for homologous antibodies by either a slide or tube agglutination test to aid in the diagnosis of Salmonellosis. The recalling firm also advised that they would issue replacements for the referenced lot number under Catalog number 240731, which would not be available until approximately July 15, 2010. Consignees were requested to return enclosed response form. |
| Quantity | 93 units |
| Distribution | Product was sold to distributors nationwide and to affiliates in Belgium and Korea. |
| FDA record | Open on fda.gov |