ClearanceBrief
Recall Z-1991-2010 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

Boston Scientific Corporation

FLEXIMA Biliary Stent System, 11.5Fr-7cm, M00539360, Sterile EO, Boston Scientific, Spencer, IN. The Flexima Biliary Stent System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.

DateAug 25, 2010 (initiated Apr 27, 2010)
StatusTerminated
ReasonFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Root causePackaging
ActionBoston Scientific sent an "Urgent Medical Device Recall - Immediate Action Required" letter date April 27, 2010 to all consignees. The letter included instructions to immediately discontinue use and segregate the affected product for return, and complete and return by fax a Reply Verification Tracking Form. Customers with affected product were instructed to contact customer service at 866-868-4004 to obtain a Return Authorization (RGA) number, and package and ship the product to Boston Scientific.
Quantity15
DistributionWorldwide Distribution: United States and Puerto Rico, and countries of Arab Emirates, Argentina, Austria, Belgium, Canada, China, Columbia, Croatia, El Salvador, Estonia, Finland, France, Great Britain, Germany, Greece, Hungary, Iran, Israel, Italy, Japan, Korea, Latvia, Lebanon, Malaysia, Mexico,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K965147BILIARY STENT, BILIARY STENT SYSTEM, BILIARY ENDOPROSTHESIS
Boston Scientific Corp, Jun 30, 1997