Recall Z-1897-2010 ยท HLF Device, Measuring, Lens Radius, Ophthalmic

Haag-Streit USA Inc

Lenstar LS 900, Biomicroscope, Slit-Lamp, AC-powered. HAAG-STREIT AG, Switzerland. Used for obtaining ocular measurements and performing calculations to assist in determination of the appropriate power and type of Intraocular Lens for implantation after removal of the natural crystalline lens following cataract removal.

DateJun 24, 2010 (initiated May 19, 2010)
StatusTerminated
ReasonDuring the creation of the Lenstar LS 900 software, the ability to perform a manual adjustment of the retinal thickness measurement was not disabled for products targeted for US distribution. The current software installed allows the user to manually alter the retinal thickness measurement, which could modify the automatic axial eye length measurement resulting in the selection of an incorrect in
Root causePMA
ActionHAAG - Streit USA, Inc. issued notification letters to their customers informing them of the problem. They instructed their customer to refrain from adjusting the gates for the measurement of retinal thickness until after the system has been updated. For further information, contact HAAG - Streit USA, Inc. at 1-513-336-7255.
Quantity45 units
DistributionNationwide Distribution -- Including AL, AZ, CA, FL, ID, KY, LA, MA, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, VA, VT, WA and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082891LENSTAR, MODEL LS900
Haag-Streit AG, Oct 20, 2009