Recall Z-1941-2010 ยท FQO Table, Operating-Room, Ac-Powered

Steris Corporation

STERIS 4085, 5085 & 5085SRT Surgical Tables

DateJun 30, 2010 (initiated May 14, 2010)
StatusTerminated
ReasonSTERIS has learned that a limited quantity of manifold assemblies were manufactured by our supplier with incorrect washers and springs in the valves of the manifold. This supplier error has the potential to cause: a) an inability to articulate/move a table section when commanded by the operator; b) two different articulations occurring when a single articulation is commanded, or c) that the table
Root causeNonconforming Material/Component
ActionOn May 14, 2010 the firm sent FCL letters to their customers notifying them of the problem. They informed their customers that a service representative will contact them to arrange correction of the unit.
Quantity79 units
DistributionWorldwide distribution: USA, Australia, India, Korea & New Zealand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090136STERIS CMAX XLT SURGICAL TABLE
STERIS Corporation, Mar 20, 2009