ClearanceBrief
Recall Z-1870-2010 ยท ESW Prosthesis, Esophageal

Boston Scientific Corporation

Ultraflex Esophageal Stent System. Covered - Large - Distal Release - 23/28/10/7. Catalog Number M00514200. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Intended for maintaining esophageal luminal patency in esophageal structures caused by intrinsic and/or extrinsic malignant tumors only and occlusion of concurrent esophageal fistulas.

DateJun 25, 2010 (initiated May 17, 2010)
StatusTerminated
ReasonSuture related device failures during deployment and repositioning of the stent
Root causeDevice Design
ActionBoston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
DistributionWorldwide Distribution - Only US distributed product is affected by recall
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032930ULTRAFLEX ESOPHAGEAL NG STENT SYSTEM
Boston Scientific Corp, Oct 23, 2003