Recall Z-1971-2010 ยท LGC Enzyme Linked Immunoabsorbent Assay, Herpes Simplex Virus, Non-Specific

Biokit U.S.A. Inc.

Sure-Vue HSV-2 Rapid Test, Catalogue No: 23-046395. 20 test Kit In Vitro Diagnostic test for Herpes Simplex Virus 2

DateJul 8, 2010 (initiated Apr 14, 2010)
StatusTerminated
ReasonNegative serum was giving a positive result Herpes Simplex Type 2 (HSV-2). It is possible that misdiagnosis (i.e., positive) could be made.
Root causeNonconforming Material/Component
ActionBiokit USA notified US Customers by fax on 2/18/10 of the reason for the kit recall and requested to discard any remaining units. Accounts were advised to retest patients or review test results. Questions are directed to the firm at 800-926-3353.
Quantity8 kits
DistributionUS (CA, FL, GA, IN, KY, MA, NY, VA) and UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983886POCKIT HSV 2 RAPID TEST