Recall Z-1886-2010 ยท LJT Port & Catheter, Implanted, Subcutaneous, Intravascular

Navilyst Medical, Inc

Navilyst Medical, Inc., Vaxcel(TM) Port Plastic Standard Port with 8F Silicone Catheter (1.5 mm/2.6 mm), REF Catalog No. 45-340, UPN Product No. M001453400, STERILE, Rx ONLY, For single use only --- Implantable Vascular Access System --- (kit contains Huber needle) -- previously distributed by Boston Scientific Corporation.

DateJul 1, 2010 (initiated May 13, 2010)
StatusTerminated
ReasonFDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.
Root causeNonconforming Material/Component
ActionNavilyst Medical, Inc. is recalling Vaxcel Ports and Vaxcel Ports with PASV Valve with valid expiration dates worldwide. Navilyst sent "Urgent Medical Device Recall - Immediate Action Required" letter/recall notification packages on May 13, 2010, via Federal Express to 471 domestic consignees; on May 14, 2010, via e-mail to 14 international consignees and to Boston Scientific Corporation (prior owner of the affected products) via facsimile and Federal Express. On June 3, 2010, Navilyst Medical sent Recall Notification Packages on behalf of Boston Scientific Corp. via Federal Express to 182 d
DistributionWorldwide distribution:USA and countries of Belgium, Brazil, Canada, Colombia, Croatia, Germany, Greece, Hong Kong, Italy, Saudi Arabia, South Korea, Spain, Tunisia, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K903364GERARD MEDICAL, INC. TRIM-PORT
Gerard Medical Enterprises, Inc., Oct 31, 1990
K942623TRIMPORT
Gerard Medical Enterprises, Inc., Aug 23, 1994
K945250TRIMPORT
Gerard Medical Enterprises, Inc., Mar 7, 1995
K954258TRIMPORT IMPLANTABLE ACCESS SYSTEM (MODIFICATION)
Gerard Medical Enterprises, Inc., May 24, 1996