ClearanceBrief
Recall Z-1876-2010 ยท LJT Port & Catheter, Implanted, Subcutaneous, Intravascular

Navilyst Medical, Inc

Navilyst Medical, Inc., Vaxcel(TM) Port with PASV(TM) Titanium Mini-Port with PASV Valve and 8F Silicone Catheter (1.4 mm/2.6 mm), REF Catalog No. 45-220, UPN Product No. M001452200, STERILE, Rx ONLY, For single use only --- Implantable Vascular Access System with PASV(TM) Valve Technology --- (kit contains Huber needle) -- previously distributed by Boston Scientific Corporation.

DateJul 1, 2010 (initiated May 13, 2010)
StatusTerminated
ReasonFDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.
Root causeNonconforming Material/Component
ActionNavilyst Medical, Inc. is recalling Vaxcel Ports and Vaxcel Ports with PASV Valve with valid expiration dates worldwide. Navilyst sent "Urgent Medical Device Recall - Immediate Action Required" letter/recall notification packages on May 13, 2010, via Federal Express to 471 domestic consignees; on May 14, 2010, via e-mail to 14 international consignees and to Boston Scientific Corporation (prior owner of the affected products) via facsimile and Federal Express. On June 3, 2010, Navilyst Medical sent Recall Notification Packages on behalf of Boston Scientific Corp. via Federal Express to 182 d
DistributionWorldwide distribution:USA and countries of Belgium, Brazil, Canada, Colombia, Croatia, Germany, Greece, Hong Kong, Italy, Saudi Arabia, South Korea, Spain, Tunisia, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030083VAXCEL WITH PASV PORT
Boston Scientific Corp, Jan 28, 2003
K032008VAXCEL WITH PASV PORT, MODEL 45-213
Boston Scientific Corp, Jul 14, 2003