Recall Z-2007-2010 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Oridion Medical 1987 Ltd.
Microstream SMART CAPNOLINE PLUS O2 100UN Part Number: 010210 Oridion 3012176-112 Physio Control
| Date | Jul 13, 2010 (initiated May 26, 2010) |
| Status | Terminated |
| Reason | The Microstream Filterline may not be recognized by the monitor when connected. No CO2 readings will be displayed and no alarms may sound for unmonitored patients. Replacing the filterline may delay treatment. |
| Root cause | Process design |
| Action | Oridion notified customers by letter on 5/26/10 and requested that they remove the FilterLines from the related production lots from the field and an immediate replacement with FilterLines from subsequent production lots. Questions are directed to the responsible person at the supplier named in the letter, or to Oridion Capnography at 888-674-3466. |
| Quantity | 1119 boxes |
| Distribution | AZ, CA, FL, HI, LA, , MA, MO, NJ, OH, PA, WA, WI, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K011536 | MICROSTREAM O2/CO2 ORAL NASAL FILTERLINE Oridion Medical 1987 , Ltd., Nov 14, 2001 |