ClearanceBrief
Recall Z-2000-2010 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Oridion Medical 1987 Ltd.

Microstream NIVLINE ADULT 25UN Part Number: 008174

DateJul 13, 2010 (initiated May 26, 2010)
StatusTerminated
ReasonThe Microstream Filterline may not be recognized by the monitor when connected. No CO2 readings will be displayed and no alarms may sound for unmonitored patients. Replacing the filterline may delay treatment.
Root causeProcess design
ActionOridion notified customers by letter on 5/26/10 and requested that they remove the FilterLines from the related production lots from the field and an immediate replacement with FilterLines from subsequent production lots. Questions are directed to the responsible person at the supplier named in the letter, or to Oridion Capnography at 888-674-3466.
Quantity4275 units (171 boxes/25 units)
DistributionAZ, CA, FL, HI, LA, , MA, MO, NJ, OH, PA, WA, WI,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980325MICROSTREAM NASAL CANNULA FILTERLINE
Oridion Medical , Ltd., Apr 16, 1998