Recall Z-0473-2011 ยท MKJ Automated External Defibrillators (Non-Wearable)
Philips Medical Systems
HeartStart FR2+ automated external defibrillator (AED) model: M3841: LAERDAL brand; text prompts only on screen (NO ECG); no configurable manual charge.
| Date | Nov 26, 2010 (initiated May 28, 2010) |
| Status | Terminated |
| Reason | Failure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable. |
| Root cause | Nonconforming Material/Component |
| Action | Philips Medical Systems issued an Urgent Medical Device Recall letter dated May 28, 2010 identifying the affected device and actions to be taken. Customers were instructed to review their inventory for affected devices. If affected devices are found, the customer should contact Philips to arrange for a replacement device. The customer should also run a battery insertion test. If the device passes the test, the AED may remain in service until a replacement is received. For further information, contact Philips at 1 800 263-3342, option 6. |
| Quantity | 1,478 units |
| Distribution | Worldwide Distribution: Throughout the USA, and to the following countries: ALBANIA, ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BERMUDA, BOSNIA AND HERZEGOVINA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, FINLAND, FRANCE, GERMANY, HONG KONG, ICELAND, INDIA, INDONESIA, IRELAND, I |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K003565 | HEARTSTREAM FR2 AED, MODEL M3860A,M3861A Agilent Technologies, Inc., Dec 20, 2000 |
| K003819 | HEARTSTREAM ATTENUATED DEFIBRILLATION PADS, MODEL M3870A Agilent Technologies, Inc., May 2, 2001 |
| K013425 | HEARTSTREAM/HEARTSTART FR2 AED ECG CABLE, MODELS M3860A, M3861A, M3840A, M3841A Philips Medical Systems, Jan 14, 2002 |
| K014157 | HEARTSTREAMFR2 AED WITH M3848A AND M3849A, MODELS M3860A, M3861, M3840A, M3841A Philips Medical Systems, Jan 17, 2002 |
| K051632 | PHILIPS HEARTSTART FR2+ AUTOMATED EXTERNAL DEFIBRILLATOR, MODELS M3860A, M3861A, M3840A, M3841A Philips Medical Systems, Oct 5, 2005 |