ClearanceBrief
Recall Z-1952-2010 ยท MBI Fastener, Fixation, Nondegradable, Soft Tissue

Biomet, Inc.

Artificial ligament fixation device # 7PE ZIPLOOP Extended Toggle, Sterile, REF 904755, Biomet Sports Medicine Warsaw, IN. Intended for soft tissue to bone fixation.

DateJul 1, 2010 (initiated Mar 31, 2010)
StatusTerminated
ReasonThe firm has received complaints that the saddle is too long. The saddle is supposed to be between 7/8 and 1 1/8 inches long. The saddle in the complaints was up to 1.9 inches long.
Root causeProcess control
ActionBiomet Sports Medicine issued an "Urgent Medical Device Recall Notice" dated March 31, 2010 to user facilities and distributors. A second notification containing updated information was issued June 2, 2010. Consignees were instructed to discontinue use of the product and return it to the firm along with a completed Response Form. For further information, contact Biomet at 1-574-371-3755 or 1-574-371-3756, Monday through Friday, 8AM to 5PM.
Quantity2466 both products
DistributionWorldwide Distribution -- United States, Canada, Australia, The Netherlands, Belgium, Italy, France, New Zealand, Austria, United Kingdom, Chile, Greece, Poland, Denmark, Turkey, Portugal, South Africa and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083070TOGGLELOC SYSTEM
Biomet Sports Medicine, Dec 16, 2008